Draft Data Validation Plan Review Form Page 13

ADVERTISEMENT

Draft Data Validation PRF-Tier II
Date: March 2, 2004
Page 13
Revision: 4.0
1.0
Semi-Volatile Data Review - Initial Calibration
Initial Calibration
Semi-volatile organic methods rely on an Initial Calibration in which 5-9 different concentrations of all target analytes are
analyzed to establish a “working range” for sample analysis. The results of this initial calibration may be used for a period of
weeks in which the initial calibration will be confirmed by Continuing Calibration Verification (CCV) Quality Control (QC)
samples. Problems with the Gas Chromatograph (GC) column (e.g., organic “junk” accumulating in the end) may require that
the end of the column be clipped off. As the retention time data is modified from this action, the need for a new Initial
Calibration must be triggered. Samples from this point on may all refer back to this new initial calibration date.
1.1
Is the Gas Chromatograph/Mass Spectrometer (GC/MS)
Indicate samples affected and qualification of sample results:
system tuned to meet the criteria for 50 ng injection of
Decafluorotriphenylphosphine (DFTPP) prior to sample
analysis?
Note: The ion abundance criteria used in this checklist is
from SW-846, Method 8270C. U.S. EPA’s National Functional
Guidelines have slightly different criteria.
Mass
Ion Abundance Criteria
51
30% to 80% of mass 198
68
<2% of mass 69
70
<2% of mass 69
127
25% to 75% of mass 198
197
<1% of mass 198
198
Base peak, 100% relative abundance
199
5% to 9% of mass 198
275
10% to 30% of mass 198
365
>0.75% of mass 198
441
Present but less than mass 443
442
>40-110% of mass 198
443
15% to 24% of mass 442
Note: All ion abundances must be normalized to m/z 198
(except the ones that are normalized to m/z 69, as noted
above).
Action: If a base peak, other than m/z 198 is used for
normalization, then qualify all associated data as “R.”
If the ion abundance criteria is not met then judgement should be
used to qualify associated data as “J” or “R.”
1.2
Has the GC/MS system hardware been tuned to meet
DFTPP tuning criteria within 12 hours prior to sample
analysis?
Note: Tuning is valid up to 12 hours after tuning analysis.
Assess whether the laboratory tuned the instrument at the
required time(s). If necessary, obtain a run log chart from the
laboratory to confirm the time of tuning.
Action: If the tune was not performed at the required time, the
Data Validator may qualify all data as “R” or “J.”

ADVERTISEMENT

00 votes

Related Articles

Related forms

Related Categories

Parent category: Business