Date: March 2, 2004
Page 20
Revision: 4.0
4.0
Semi-Volatile Data Review - Target Compound Identification
Target Compound Identification
One of the primary goals of sample analyses is to accurately identify the presence of target compounds. The procedure used
by SW-846, Method 8270C to identify the presence of target compounds is to compare the retention time of unknown
compounds in a sample to the retention time of knowncompounds in the calibration standard. The Relative Retention Time
(RRT) of the compound in the sample must be within +/- 0.06 units of the RRT of the initial calibration standards. In addition,
the mass spectra of the standard must match the mass spectra of the calibration standard within certain tolerances . For
example, all ions in the standard mass spectra at a relative intensity of 10% must also be present in the sample spectrum.
In addition, the relative intensities of the ions in the sample must agree within +/- 20% between the sample and the standard
spectra. Ions that are present at greater than 10% in the sample mass spectrum, but not present in the standard spectrum
must be considered and accounted for. Results for unknown compounds in samples are compared to an electronic ”library”
of compounds to determine if all results (retention time and mass spectra) match within specified tolerances to a library
compound (or at least to a certain type of compound of specific size).
4.1
Are the RRTs of reported compounds in the ICV within
+/- 0.06 RRT units (minutes) of the RRT for the initial
calibration?
Action: Use this information in conjunction with the mass spectral
analysis information to qualify data. If target compound
identification is in error, qualify that compound as “R.”
4.2
Check the sample compound spectra against spectral
data of the standard. All ions in the standard mass
spectra at a relative intensity of 10% must also be in the
identified sample compound.
Action: Use this information in conjunction with the mass spectral
analysis information to qualify data. If target compound
identification is in error, qualify that compound as “R.”
4.3
Check the sample compound spectra against spectral
data of the standard. The relative intensity of the ions in
the sample must agree within +/- 20% of the value
obtained from the standard spectrum.
Action: Use this information in conjunction with the mass spectral
analysis information to qualify data. If target compound
identification is in error, qualify that compound as “R.”