Form Cms-1882 - Portable Xray Survey Report Page 8

ADVERTISEMENT

name of facility
not
standards
code
met
n/a
eXplanatory statement
met
list and indicate for each piece of equipment
total
Half-value
name or
kvp
serial number
filtration
layer
H45
(c) termination of exposure
A device is provided to terminate the exposure after a preset
time or exposure.
H46
(d) control panel
the control panel provides a device (usually a milliammeter or
a means for an audible signal) to give positive indication of the
production of X-rays whenever the X-ray tube is energized.
the control panel includes appropriate indicators (labelled
control settings and/or meters) which show the physical factors
(such as kVp, mA, exposure time or whether timing is
automatic) used for the exposure.
H47
(e) exposure control switch
The exposure control switch is of the dead-man type and is so
arranged that the operator can stand at least 6 feet from the
patient and well away from the useful beam.
H48
(f) protection against electrical hazards
only shockproof equipment is used. all electrical equipment is
grounded.
H49
(g) mechanical supporting or restraining devices
mechanical supporting or restraining devices are provided so
that such devices can be used when a patient must be held in
position for radiography.
8
form cms-1882 (02/09)